INSTALLATION
28
3.3 ELECTRICAL CONNECTION
The CG 4 is provided with a 10 m, 6 wire shielded cable (8 wire for the
Pt-100 version). The following colour code is used:
Red : Plus thermopile
Blue : Minus thermopile
White : Case
Black : Shield
Thermistor
Yellow
Green
Pt-100 (optional)
Yellow : Pt 100 (combined with brown)
Brown : Pt 100 (combined with yellow)
Green : Pt 100 (combined with grey)
Grey : Pt 100 (combined with green)
A surge arrester is installed to lead off induced lightning currents to
the case. It is recommended to ground the case for this reason. The
surge arrester is noble gas filled, has infinite impedance and recovers
after breakdown. Breakdown voltage is 90 V. Peak pulse current is 10
kA.
The shield is isolated from the case, so no shield-current can exist.
Shield and white lead may be connected to the same ground at the
readout equipment. The cable must be firmly secured to minimise
spurious response during stormy weather (deforming standard cable
produces voltage spikes, a tribo electric effect and capacitance effect).
Kipp & Zonen pyrgeometer cables are of low noise type, however take
care that the terminals '+' and '-' at a connection box have the same
temperature, to prevent thermal EMF's.
A junction box or connector with a metal outer case is advised.